Résumé

Based on a radiation field produced by an electrostatic accelerator, a radiation survey meter test and calibration facility has been designed and characterized in terms of air kerma and ambient dose equivalent. The electron beam impinges a tantalum target to produce X-rays. The spectrum has been measured and calculated. Traceability to the International System of units is achieved by means of a calibration with a primary dosimeter for air kerma.
Fichier principal
Vignette du fichier
papier CIM JM bordy V5.pdf (1005.11 Ko) Télécharger le fichier Fichier principal
Vignette du fichier
presentation JMB.pdf (1.96 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

cea-04557978 , version 1 (24-04-2024)

Identifiants

Citer

jean-marc bordy, Valentin Blideanu, Arnaud Chapon, Gabriel Dupont, Dorin Dusciac, et al.. Primary reference in terms of air kerma for ionizing radiation fields produced by an electrostatic electron accelerator. CIM2019 - 19th International Congress of Metrology, Sep 2019, Paris, France. pp.15003, ⟨10.1051/metrology/201915003⟩. ⟨cea-04557978⟩